Health

FDA flags Tandem Mobi insulin pump recall as most serious

Some Tandem Mobi pumps can sound false blockage alarms that pause insulin delivery. Tandem has reported 1,248 serious injuries and no deaths.

File photo: The FDA's Center for Devices and Radiological Health in Silver Spring, Md.
File photo: U.S. Food and Drug Administration

FACTS

The U.S. Food and Drug Administration said on Wednesday, Nov. 25, that Tandem Diabetes Care’s Tandem Mobi insulin pump recall is the most serious type of medical device recall, according to the FDA’s recall notice. Certain pumps can sound false blockage alarms and pause insulin delivery, which can raise blood sugar. Tandem has reported 1,248 serious injuries and no deaths as of Oct. 28, the FDA said.

What the insulin pump recall covers

The recall applies to all Tandem Mobi Insulin Pumps with Interoperable Technology, the FDA said. Pumps whose serial numbers appear on a list of affected products must be returned and replaced, while other pumps may stay in use with updated instructions, according to the notice. Tandem sent letters to customers on Sept. 23, 2026, and also notified health care providers, U.S. distributors and pharmacies, the agency said.

According to the FDA, certain pumps may generate false occlusion alarms, which warn of a blockage in insulin flow, “when no actual insulin flow obstruction is present.” The pump may then suspend insulin delivery until the user acknowledges the alarm and resumes therapy. Repeated interruptions may raise blood glucose, and severe cases may require medical intervention or hospitalization, the FDA said.

How many pumps and what Tandem says caused it

Tandem began the recall in late September, and the FDA’s recall database entry is dated Oct. 30, MedTech Dive reported on Nov. 11. The recall covers more than 80,900 Mobi devices, mostly distributed in the United States, the outlet reported.

Tandem attributes the false alarms to increased friction in the pump’s drive module, which can reduce motor power and drop motor speed below the threshold the pump uses to detect a blockage, according to MedTech Dive. The outlet reported that Tandem said it is investigating the issue and has taken steps during testing to keep it from happening again.

In that Nov. 11 report, MedTech Dive cited a Tandem spokesperson as saying the company had received 230 reports requiring medical intervention or hospitalization as of Oct. 29. The FDA’s Nov. 25 notice puts the number of serious injuries reported by Tandem at 1,248 as of Oct. 28. Neither source explains the difference between the two figures.

MedTech Dive reported that this recall is separate from an earlier Tandem Mobi recall that the FDA posted in April 2026. That one involved a software malfunction that could affect insulin delivery, and Tandem asked users to install an update, the outlet reported.

What the Tandem Mobi is

Tandem describes the Mobi on its products page as the “World’s smallest, durable AID system,” a reference to automated insulin delivery, and says it is about half the size of the company’s t:slim X2 pump. The company says users can control the pump fully from a compatible smartphone and that it runs Control-IQ+ software, which can raise, lower or stop basal insulin based on predicted glucose levels. Tandem says the pump is indicated for people 2 years of age and older.

What the FDA tells users to do

For pumps on the replacement list, the FDA says users should submit a return and replacement request right away, through the link in Tandem’s letter or by calling the company’s customer technical support. Until then, the agency says, they should keep push notifications turned on in the Tandem Mobi app, respond to alarms as the User Guide describes, check blood glucose regularly and keep a backup method of taking insulin ready.

Users whose pumps are not on the list may keep using them, the FDA said. They should follow the occlusion alarm troubleshooting steps, turn on app push notifications, keep a backup insulin method ready and contact Tandem customer support if alarms become unusually frequent or keep disrupting therapy. The guidance also includes checking the cartridge, tubing and infusion site for blockages or damage, RTTNews reported. The pump sounds increasingly loud alerts or vibrates until an alarm is acknowledged, according to RTTNews.

The FDA asks patients and health care providers to report problems with the device to its MedWatch program. Tandem’s support contacts are listed in the FDA notice.

More health news is on the Stay Current News Health page.